Program-Controlled Leakage Current Tester for Testing Medical Equipment
1. Introduction
RK7500Y series medical leakage current tester is a programmable and one-key serialization test equipment, built-in medical simulation human body network, it can perform ground leakage current, touch current, patient leakage current, patient auxiliary current and etc. tests as per IEC60601-1 and IEC61010-1.
2. Features
1). Built-in medical simulation human body network
2). It can synchronize complete the test of ground leakage current, touch current, patient leakage current (DC and AC) and others, or the tests can be performed separately.
3). Test items can be set according to categories, and the same test item can be automatically combined for testing, and simultaneously display the measured voltage value, external voltage value, normal leakage current, and single fault state leakage current value.
4). Leakage current range: support maximum 10mA leakage current RMS range
5). LCD2004C liquid crystal display
6). With PLC interface, it can be controlled externally; with RS232C interface, it can upload test data in real time
3. Parameters
Model | RK7505Y | RK7510Y | RK7520Y | RK7530Y | RK7550Y | ||||
Test device | Automatic range conversion, true RMS measurement | ||||||||
IEC60601 standard |
Leakage current measurement range(AC): I (3~99.9)uA resolution: 0.1uA II (100.0~999.9)uA resolution: 0.1uA III (1000~9999) uA resolution: 1uA Patient current, patient auxiliary current measurement range (DC): (3~999.9)uA resolution: 1uA Accuracy: ±5%+3 digits (>10uA) Frequency response range: DC~1MHz Measuring Impedance Circuit: Meet the requirements of IEC60601 figure 12 |
||||||||
IEC61010 standard |
Leakage current measurement range(AC): I (3~99.9)uA resolution: 0.1uA II (100.0~999.9)uA resolution: 0.1uA III (1000~9999) uA resolution: 1uA Accuracy: ±5%+3 digits(>50uA) Measuring Impedance Circuit (MD): Meet the requirements of IEC60601 figure A.1 |
||||||||
Output Voltage | Measuring the voltage output range of the power supply: 0V-300V, resolution: 1V, accuracy: ±5%+2 digits | ||||||||
output voltage capacity | 500VA | 1000VA | 2000VA | 3000VA | 5000VA | ||||
Current upper limit setting | Range: (3-9999)uA, resolution: 1uA, accuracy: ±4%+3 digits (>10uA) | ||||||||
Timing device | Range: (3-9999)s, resolution: 1s, accuracy: ±5% | ||||||||
Working condition | 0-40℃, ≤75%RH | ||||||||
Working power | 220V±10%, 50Hz/60Hz±3Hz | ||||||||
Interface | RS232, PLC | ||||||||
No. of memory group | 5 groups | ||||||||
Screen | LCD2004c | ||||||||
Dimension(D*H*W) | 445*160*352mm | 445*160*352mm | 533*210*430mm | 533*210*430mm | 610*243.5*430mm | ||||
Weight | 16kg | 23.66kg | 38.7kg | 49.2kg | 74.5kg | ||||
Power comsuption | <50W | ||||||||
Accessories | Power cable, RS232 cable, RS232 to USB cable, test wire |