Add to Cart
The Measuring Circuits for Touch Current for Electrical Burns at High Frequencies is designed and manufactured according to the requirements of IEC60990 figure 3 and IEC61010-1 clause 6.3.1 and Figure A.3.
IEC60990 Methods of measurement of touch current and protective conductor current Figure 3
IEC61010-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance Clause 6.3.1 and Figure A.2
IEC60990 | IEC61010-1 | Parameters |
RS | R1 | 1500Ω±5% |
RB | R2 | 500Ω±5% |
CS | C1 | 0.22μF±10% |
Voltage meter | Voltage meter | Provided by user |